Myron L – Ultrapen – PT1, COND., TDS, Salt, Temp

This instrument is designed to be extremely accurate, fast and simple to use in diverse water quality applications. Advanced features include the ability to select from 3 different solution types that model the characteristics of the most commonly encountered types of water; proprietary temperature compensation and TDS conversion algorithms; highly stable microprocessor-based circuitry; user-intuitive design; and waterproof housing. A true, one-handed instrument, the PT1 is easy to calibrate and easy to use. To take a measurement, you simply press a button then dip the pen in solution. Results display in seconds.

HIGH PERFORMANCE FEATURES                               

Accuracy of ±1% of READING ±.2% at calibration point

Reliable Repeatable Results

Automatic temperature compensation

Autoranging

Accuracy of benchtop lab equipment in an affordable, portable, compact pen

Rugged, reliable, shock-resistant fully encapsulated electronics, body constructed of aircraft aluminum

Simple one button function

Specially designed with measurement ranges and solution modes to meet the water quality testing needs of many applications

High capacity N-type battery included

Waterproof

APPLICATIONS                                                       

Agriculture

Hydroponics

Desalination

Aquaculture

Pharmaceutical

Boilers and Cooling Towers

Fountain Solutions for Printing

Horticulture

Wastewater

Pool and Spa

Food and Beverage

Irrigation

Reverse Osmosis

Please call for questions or technical assistance! 714-564-9100

Measurement Range: 1 – 9999 μS or ppm (0.0010 – 9.999 ppt salinity)
Accuracy: ± 1% of reading
Repeatability: <1000 μS or ppm ± 1 Count≥ 1000 μS or ppm ± 0.3% of reading
Resolution: Electrical Conductivity (EC) and TDS:0.1 for 1.0 – 99.9 μS or ppm1 for 100 – 9999 μS or ppmSalinity:0.0001 for 0.0010 – 0.0999 ppt0.001 for 0.100 – 9.999 pptTemperature:0.1 ºC or ºF
Time to Reading Stabilization: 10 – 20 seconds
Active Mode Power Consumption: 30 – 100 mA
Sleep Mode Power Consumption: 2 μA
Temperature Measurement Range: 0 – 71 °C or 32 – 160 °F
Temperature Accuracy Displayed: ± 0.1 ºC or ± 0.2 ºF
Temperature Compensation Method: Automatic to 25 ºC
Physical Dimensions: 17.15 cm L x 1.59 cm D or 6.75 in. L x .625 in. D
Weight: 55 g or 1.94 oz.
Case Material: Anodized Aluminum with UV Protective Coating
Battery Type: N type, Alkaline
Battery Voltage: 1.5 V
Calibration Solution Point: 1800 μS KCl; 3000 ppm 442™ (2027 ppm NaCl)
Operating/Storage Temperature: 0 – 55 ºC or 32 – 131 ºF
Water Resistance: IP67 and NEMA 6
EN61236-1: 2006 – Annex A: 2008 Electrostatic discharge to case of instrument maycause PT1 to spontaneously power on. In this case,the PT1 will power off after several seconds.